A Coriolis meter does not measure flow. It measures how late one side of a shaking tube is. From that lateness it reads mass flow directly, which is usually the thing you wanted to know. Move the flow slider in the model below and watch the lateness change.
Watch the physics
Press Tour to see the five stages, or use the numbered buttons. Under stage 3 you can drag the flow slider from negative to positive and watch the time lag follow it.
The narrated walkthrough
The full version, about two and a half minutes, narrated with a synthetic voice. It uses the same numbers as the model.
Shake, push, lag
The model is one illustrative water meter with a U-shaped tube. Every figure below is invented for the example and recomputed by a second program.
- Shake. A coil drives the tube at its natural frequency. Empty, the tube rings at 180.0 Hz. Fill it with water and the extra mass (0.2495 kg on top of the tube’s 0.300 kg) lowers the pitch to 133.0 Hz, because f = (1 ÷ 2π) × √(k ÷ m).
- Push. Let water flow at 2.00 kg/s. On the inlet leg the fluid is pushed outward along a swinging arm, so it drags back on the tube. On the outlet leg it is heading home and pushes the tube on. The Coriolis force is F = 2 · ṁ · Ω, where ṁ is the mass flow and Ω is how fast the leg is rotating. With a 1.0 mm tip swing on a 0.35 m leg, Ω peaks at 2π × 133.0 × 0.00286 = 2.39 rad/s, so the force is 2 × 2.00 × 2.39 = 9.6 N per metre of leg, or 3.3 N on the leg, in opposite directions on the two legs.
- Lag. Equal and opposite pushes twist the tube, so one side reaches the middle of its swing slightly after the other. In the model, 2.00 kg/s gives 40.0 µs between the two pickoff signals. Double the flow and the lag doubles to 80.0 µs. Stop the flow and it vanishes. Reverse it and the lag changes sign, so the meter also knows the direction. Mass flow is the lag times a calibration factor, here 0.05 kg/s per µs.
The lag is tiny. At 133.0 Hz a 40.0 µs delay is a phase difference of only 1.9 degrees, which is why the animation slows the vibration by about 200 times and draws the gap between the legs 16 times larger. The scope in the model is not exaggerated sideways: it zooms in on one zero crossing, so the gap you see is the true 40.0 µs.
The same tube reads density
Heavier fluid, lower pitch. Rearranging the frequency formula gives the density: ρ = (m ÷ V) × ((f₀ ÷ f)² − 1). With the tube mass 0.300 kg, internal volume 0.25 L and empty frequency 180.0 Hz, a pitch of 133.0 Hz reads 998 kg/m³ (water) and 137.7 Hz reads 850 kg/m³ (an oil).
That is why the meter is a mass meter first. The same 40.0 µs lag means 2.00 kg/s for either fluid, but the volume flow is 2.00 L/s of water and 2.35 L/s of oil, because volume flow is mass flow divided by density. You get mass without ever correcting for what the fluid is.
In the field, watch the density
Gas in the liquid is the classic trouble. In the model, 5% gas by volume pulls the density reading from 998 down to 948 kg/m³ and nudges the pitch to 134.5 Hz, while the pump keeps running. The density shows it before anything else does, which makes a steady flow with a falling density a warning. Sources also point to zero offset at no flow, pipe vibration, and a coated or corroded tube as things to check.
What this model leaves out
It is a teaching model, not a design. Real meters have several tube shapes, not only the U. They compensate for temperature, pressure and the exact drive frequency, and they calibrate the density relation against reference fluids rather than a single formula. Real tolerance to gas, and the right sign convention for which pickoff leads, depend on the make and model, so read a manufacturer’s manual before relying on any detail here. Nothing in the model is a claim about a particular product or site.
The question to ask about any Coriolis reading
A flow number alone cannot tell you whether the meter is healthy. The density that comes with it can.
If a Coriolis meter on your plant showed a steady flow but its density reading fell by 5% in a minute, what would you check first?